Measure Frequency on CH1
Figure
4
-
8: Measure Current on CH2 with Internal Loop Power (Range: 55
mA)
4.7 Measure Frequency on CH1
Figure
4
-
9: Example A – Measure Frequency on CH1 (Range 0 to 50
kHz)
1. Set applicable channel options: CH1, Source, Frequency, Hz (or other unit).
2. Complete the electrical connections.
3. Default channel settings:
•
Range: 0 to 50
kHz
•
Trigger level: 2.5
V
If necessary, adjust the Trigger level value by pressing the SETTINGS
icon and
selecting MANUAL LEVEL.
4. AUTO TRIGGER can be enabled or disabled.
Note:
Manual Level setting is for Manual Trigger only.
Содержание Druck DPI 620 Genii
Страница 1: ...Druck com DPI 620 Genii Advanced Modular Calibrator Instruction Manual...
Страница 2: ......
Страница 12: ...Copyright 2013 Baker Hughes Company x DPI 620G Instruction Manual English...
Страница 20: ...Copyright 2013 Baker Hughes Company 8 DPI 620G Instruction Manual English Chapter 1 Introduction...
Страница 38: ...Copyright 2013 Baker Hughes Company 26 DPI 620G Instruction Manual English Chapter 2 Instrument Operation...
Страница 62: ...Copyright 2013 Baker Hughes Company 50 DPI 620G Instruction Manual English Chapter 4 Electrical Tasks...
Страница 72: ...Copyright 2013 Baker Hughes Company 60 DPI 620G Instruction Manual English Chapter 5 Pressure Tasks...
Страница 84: ...Copyright 2013 Baker Hughes Company 72 DPI 620G Instruction Manual English Chapter 7 Data Logging...
Страница 110: ...Copyright 2013 Baker Hughes Company 98 DPI 620G Instruction Manual English Chapter 9 HART Operations...
Страница 126: ...Copyright 2013 Baker Hughes Company 114 DPI 620G Instruction Manual English Chapter 10 FOUNDATION Fieldbus...
Страница 162: ...Copyright 2013 Baker Hughes Company 150 DPI 620G Instruction Manual English Chapter 12 Calibration Procedures...
Страница 164: ...Copyright 2013 Baker Hughes Company 152 DPI 620G Instruction Manual English Chapter 13 General Specification...
Страница 165: ......