12. Calibration Procedures
Druck can provide a calibration service that is traceable to international standards.
Druck recommends returning the instrument to the manufacturer or an approved service agent
for calibration. If using an alternative calibration facility, check that it uses the standards that follow
in this chapter.
12.1 Before Starting
Use only original parts supplied by the manufacturer. To do an accurate calibration, use the
following:
•
Calibration equipment specified in Table
1.
•
Stable temperature environment: 21
±
1°C (70
±
2°F).
Before starting a calibration procedure, it is recommended leaving the equipment in the
calibration environment for a minimum of two hours.
Before starting the calibration, check the time and date on the instrument are correct.
Table
12
-
1: Calibration Equipment Specification
Function
Calibration Equipment
a
Measure
Source
Current (CH1 or CH2)
Current (mA) calibrator.
For accuracy, see Table
Current (mA) calibrator.
For accuracy, see Table
3.
Voltage (CH1)
Voltage calibrator.
For accuracy, see Table
Voltage calibrator.
For accuracy, see Table
7.
Voltage (CH2)
Voltage calibrator.
For accuracy, see Table
–
Millivolts (CH1)
Voltage mV calibrator.
For accuracy, see Table
Voltage mV calibrator. For
accuracy, see Table
6.
Millivolts (CH2)
Voltage mV calibrator.
For accuracy, see Table
–
Millivolts TC mV (CH1)
Voltage mV calibrator.
For accuracy, see Table
14.
Voltage mV calibrator.
For accuracy, see Table
Frequency (CH1)
Signal generator.
Total error: 0.3
ppm or better
Frequency meter.
Total error: 0.3
ppm or better.
Resolution: 8
digits (minimum).
Voltage calibrator.
For accuracy, see Table
7.
Resistance (CH1)
Standard resistor (100R, 200R,
300R, 400R, 1k, 2k, 4k).
Total uncertainty: 20
ppm or
better.
An ohmmeter or an RTD
measurement system with the
specified excitation currents, refer
to Table
Cold Junction (CH1)
Calibrated K type thermocouple.
Accuracy: 50
mK for
-
5 to 28°C (23
to 82.4°F)
Accuracy: 50
mK for
-
5 to 28°C (23
to 82.4°F)
Cold Junction (CH1)
Thermocouple temperature
reference unit (0°C).
Accuracy: 30
mK
–
Содержание Druck DPI 620 Genii
Страница 1: ...Druck com DPI 620 Genii Advanced Modular Calibrator Instruction Manual...
Страница 2: ......
Страница 12: ...Copyright 2013 Baker Hughes Company x DPI 620G Instruction Manual English...
Страница 20: ...Copyright 2013 Baker Hughes Company 8 DPI 620G Instruction Manual English Chapter 1 Introduction...
Страница 38: ...Copyright 2013 Baker Hughes Company 26 DPI 620G Instruction Manual English Chapter 2 Instrument Operation...
Страница 62: ...Copyright 2013 Baker Hughes Company 50 DPI 620G Instruction Manual English Chapter 4 Electrical Tasks...
Страница 72: ...Copyright 2013 Baker Hughes Company 60 DPI 620G Instruction Manual English Chapter 5 Pressure Tasks...
Страница 84: ...Copyright 2013 Baker Hughes Company 72 DPI 620G Instruction Manual English Chapter 7 Data Logging...
Страница 110: ...Copyright 2013 Baker Hughes Company 98 DPI 620G Instruction Manual English Chapter 9 HART Operations...
Страница 126: ...Copyright 2013 Baker Hughes Company 114 DPI 620G Instruction Manual English Chapter 10 FOUNDATION Fieldbus...
Страница 162: ...Copyright 2013 Baker Hughes Company 150 DPI 620G Instruction Manual English Chapter 12 Calibration Procedures...
Страница 164: ...Copyright 2013 Baker Hughes Company 152 DPI 620G Instruction Manual English Chapter 13 General Specification...
Страница 165: ......