Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 59
Set the Pressure Module to Zero
Figure
5
-
15: Example: Leak Test Results
Note:
To set the Leak Test options, a pressure module or external pressure sensor must be
correctly installed.
5.8 Set the Pressure Module to Zero
SETTINGS
> ZERO > ZERO
Use this option to write a new zero pressure value to the pressure module in use. The sensor zero
adjustment is permitted only if the adjustment is less than 10% FS positive pressure value of the
sensor.
Note:
To make a temporary adjustment for zero, use the Tare function.
Figure
5
-
16: Pressure Module Zero Example
Содержание Druck DPI 620 Genii
Страница 1: ...Druck com DPI 620 Genii Advanced Modular Calibrator Instruction Manual...
Страница 2: ......
Страница 12: ...Copyright 2013 Baker Hughes Company x DPI 620G Instruction Manual English...
Страница 20: ...Copyright 2013 Baker Hughes Company 8 DPI 620G Instruction Manual English Chapter 1 Introduction...
Страница 38: ...Copyright 2013 Baker Hughes Company 26 DPI 620G Instruction Manual English Chapter 2 Instrument Operation...
Страница 62: ...Copyright 2013 Baker Hughes Company 50 DPI 620G Instruction Manual English Chapter 4 Electrical Tasks...
Страница 72: ...Copyright 2013 Baker Hughes Company 60 DPI 620G Instruction Manual English Chapter 5 Pressure Tasks...
Страница 84: ...Copyright 2013 Baker Hughes Company 72 DPI 620G Instruction Manual English Chapter 7 Data Logging...
Страница 110: ...Copyright 2013 Baker Hughes Company 98 DPI 620G Instruction Manual English Chapter 9 HART Operations...
Страница 126: ...Copyright 2013 Baker Hughes Company 114 DPI 620G Instruction Manual English Chapter 10 FOUNDATION Fieldbus...
Страница 162: ...Copyright 2013 Baker Hughes Company 150 DPI 620G Instruction Manual English Chapter 12 Calibration Procedures...
Страница 164: ...Copyright 2013 Baker Hughes Company 152 DPI 620G Instruction Manual English Chapter 13 General Specification...
Страница 165: ......