Fyrite INSIGHT
31
4.2 Analyzer Power On & Warm Up
1. Connect the probe and make sure that the analyzer is properly set
up per Section 3.
IMPORTANT!
DO NOT insert probe into stack before powering on
the analyzer.
2. Place the probe in an area that contains fresh air. This ensures
that the sensors will be properly zeroed during the warm-up cycle.
3. Power up the analyzer by pressing the
PWR
button for at least 1
second, or until a single beep is heard.
Observe that the analyzer’s firmware version, model and serial
numbers are briefly displayed, followed by the Warm Up screen.
Version: V2.00
Model: 24-
8252
Serial: NW1060
4. After the 60 second warm-up period is completed, the instrument
will display the Combustion Test HOLD screen.
If, however, any errors are detected during warm-up, the message
“ERRORS DETECTED” will be displayed along with a list of those
errors. In this example, the O
2
sensor is missing and the T-Stack
thermocouple is not connected.
Refer to Section 6.7 for a listing and possible remedy for the errors
displayed.
4.3 Fuel Selection
The top line of Combustion Test HOLD screen shows the fuel currently
selected.