3. Properly dispose of the old CO sensor.
4. Plug the new CO sensor into its socket.
5. Install the CO cap by aligning the arrow on the cap with the
“OPEN” marker, then twisting to the “CLOSED” marker. Reattach
tubing.
6.4 Thermocouple Replacement
Replace the probe’s thermocouple as follows:
Replacement kit contains a thermocouple assembly, two O-rings, and two
wire-splice connectors.
Tools Required:
• Small flat blade screwdriver
• Wire cutter
• Wire stripper
• Slip joint pliers
Procedure:
1. Gain access to the thermocouple connections by first removing three
screws from probe handle, and then separating the two handle
pieces.
2. Cut wires attached to old crimp connectors, leaving behind as much
of the probe’s thermocouple-connector wire as possible
Probe Handle
Screws
Fig. 6.7
3. Pull old thermocouple from probe body and discard.