5.0 CAlIBRATION & MAINTENANCE
IMPORTANT:
Before performing any calibration procedure,
ensure that fresh batteries are installed or use the optional
AC power adapter. Also, ensure that the analyzer is at room
temperature and will be sampling fresh air when powered
ON.
IMPORTANT:
To maintain accuracy as listed in the Techni
-
cal Characteristics Section of this manual, the standards
used must be at least four times as accurate as the stated
accuracy of the Fyrite
®
INSIGHT.
5.1 B-Smart
™
Sensors
The Fyrite
®
INSIGHT utilizes Bacharach’s new B-Smart
™
Sensor tech-
nology for its CO sensor. The B-Smart
™
sensor is marked with a 10-digit
calibration code that can be entered in the instrument’s memory via the
keypad, or with Fyrite user software.
Benefits of B-Smart
™
Sensors:
•
New sensors can be installed without applying gas for calibration.
•
Sensors can be pre-calibrated and installed when needed.
•
Sensors can be moved from one analyzer to another.
The B-Smart
™
sensors should be calibrated by an authorized Bacharach
Service Center every 6 to 12 months to assure that the analyzer contin-
ues to meet its published accuracy specifications. The B-Smart
™
sensors,
however, can be calibrated in the field if your facility has the necessary
equipment and qualified personnel to perform the procedures described in
the following sections of this instruction manual.
5.2 Starting a Calibration
Start any calibration by performing the following:
1. With the analyzer power off, place the unit in fresh, ambient air;
then power on the analyzer.