Fyrite INSIGHT
54
5. Wait until the Measured reading stabilizes, and then press
ENTER
to calibrate the TA-Zero Measured value to that of the Applied
value. The message “Good Calibration” should briefly appear, fol-
lowed by the Calibrate TA-Span screen.
TA-Span Procedure:
6. Set thermocouple simulator to 100 °C (212 °F), and then use the
and buttons to enter an Applied value that exactly equals the
setting of the simulator.
S1
S2
S3
Calibrate TA-Span
Measured: 211.1 °F
Applied: 212.0 °F
Press ENT
S1
S2
S3
Calibrate TA-Span
Measured: 211.1 °F
Applied: 212.0 °F
Press ENT
Measured: 98
o
C
Applied:
1
00
o
C
Press ENT
Alternatively:
Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the ther-
mometer reading.
The calibration range is from 90 to 110 °C (194 to 230 °F). An at
-
tempt to calibrate outside this range will cause the message “Bad
Calibration Wrong CAL Entry” to appear in the following step.
7. Wait until the Measured reading stabilizes, and then press
ENTER
to calibrate the TA-Span Measured value to that of the Applied
value. The message “Good Calibration” should briefly appear, fol-
lowed by the re-display of the Calibration List screen.
5.7 CO Sensor Calibration
Material required:
• Calibration kit, P/N 0024-7059
• Gas cylinder: 500 ppm CO in air, P/N 0024-0492