Fyrite INSIGHT
26
Screen for normal operation. Regular calibration periods of 6 months to 1
year are recommended.
NOTE:
The default CAL Reminder period is
Never.
Set the calibration reminder period as follows:
1. Enter the Setup Menu per Section 3.7.
2. Select
CAL Reminder Period
.
3. Select the desired time period.
4. Press
ENTER
to save selection and return to the Setup Menu or
ESC
to cancel.
NOTE:
The date and time settings must be correct to get ac
-
curate calibration reminders.
3.7.15 Inactivity Timeout
This is the selected amount of time that will elapse before the analyzer
automatically powers down due to inactivity from the user. The timeout
periods available are: None, 30 or 60 minutes. However, if the CO value
is greater than 50 ppm or the O
2
value is less than 16.0%, this feature is
disabled and power down will not occur.
NOTE:
The default set time is
None.
This selection will prevent the
analyzer from automatically powering off.
• Press ▲▼ buttons to scroll/highlight options.
• Press
(
ENTER
)
to select
or save.
•
Press
ESC
to go back.