Fyrite INSIGHT
46
2. Allow the analyzer to cycle through its 60 second warm-up period.
During warm-up, the analyzer’s operation is checked and the sen-
sors are set to the following ambient conditions:
•
Oxygen sensor spanned to 20.9%
•
CO sensor zeroed
•
Pressure sensor zeroed
Any errors detected during warm-up will be listed on the display
immediately following warm-up, as displayed below.
Correct any errors before proceeding. Refer to Section 6.7 for a
list of error messages and their meanings.
3. Press the
MENU (F2)
button to access the Main Menu.
4. Use the ▲▼ buttons to select Calibration. Press
ENTER
to dis-
play the Calibration Password screen.
5. Before calibration can begin a 4-place numeric password must be
entered. Use the ▲▼ and ◄► buttons to enter the password.
(Note that the default password is 1111)
6. Press the
ENTER
button to accept the password. If the correct
password was entered, the Calibration Menu will be displayed.
7. Use the ▲▼ buttons to select the desired sensor channel to be
calibrated, and then perform the calibration procedure for that
sensor as described in the following sections.