3-5
Chapter 3. MOUNTING AND WIRING
z
Installation procedure
CAUTION
Mount securely in order to prevent vibration. Otherwise, equipment failure
could result.
Mount the device horizontally. Do not mount it with the display facing down.
Doing so might cause measurement error or equipment failure.
MQV9020/9050/9500/0005/0010/0050/0200
Install the device with two M4 screws using the mounting holes on
the base of the device.
Unit: mm
30
15
4.5
z
Installing the separate display unit
For models with a separate display
(1) Make mounting holes in the panel according to the panel cutout
dimensions.
Unit: mm
M4 (2)
35
±0.2
17.5
±0.2
100
±0.2
90
±0.2
50
±0.2
45
±0.2
Panel cutout dimensions
(2) Mount the display and secure it by tightening the screws.