![Azbil Micro Flow MQV0002 Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/azbil/micro-flow-mqv0002/micro-flow-mqv0002_user-manual_3042628077.webp)
7-9
Chapter 7. SPECIFICATIONS
Model No.
Item
MQV0050(J, K)
MQV0200
MQV0500
Flow rate
display
Display method
7-segment LED, 4 digits
Setup resolution
Gas type and control range (P. 7-11)
Indication accuracy (at standard
temperature and differential
pressure; Q: flow rate)
(1) ±1.5 % FS ±1 digit (80 % FS < Q ≤ 100 % FS)
(2) ±1.0 % FS ±1 digit (40 % FS < Q ≤ 80 % FS)
(3) ±0.5 % FS ±1 digit (0 % FS ≤ Q ≤ 40 % FS)
Totalizing
function
Display range
0.001 to 99,999.999 m
3
0.01 to 999,999.99 m
3
0.0 to 9,999,999.9 m
3
Display resolution
0.001 m
3
0.01 m
3
0.01 m
3
Totalized count backup timing
(1) Every 0.5 m
3
(1) Every 2 m
3
(1) Every 5 m
3
(2) Every hour from the previous backup
(3) At the time the RUN key is pressed.
Analog
output
Output type
Instantaneous flow rate (PV) output or setting flow rate (SP) output (switchable)
Output scale
0 to full-scale flow rate (scale can be changed.)
Output range
0 to 5 V DC/1 to 5 V DC/0 to 20 mA/4 to 20 mA (switchable)
Max. output
7 V DC max./28 mA max. (max. output when flow rate exceeds range)
Accuracy
Total output accuracy ±0.3 % FS Indication accuracy ±0.3 % FS
External load resistance
Voltage output type: 250 kΩ min. Current output type: 300 Ω max.
Alarm/event
output
Number of outputs
Alarm output: 1, Event output: 2
Output rating
30 V DC, 30 mA max. (open collector non-insulated output)
Totalized pulse output width
100 ms ±10 % (when pulse output is selected in the function setup)
Totalized pulse output rate
0.001 m
3
/1 pulse
0.01 m
3
/1 pulse
0.01 m
3
/1 pulse
External
switching
input
Input type, number of inputs
External 3-way switching input (OPEN/GND/5V): 1
External contact input (2-way switching): 3
Required circuit type
Non-voltage contact, or open collector
Contact OFF terminal voltage
External 3-way switching input: 2.5 ±0.5 V
External contact input: 2.8 ±0.5 V
Contact ON terminal current
Approx. 0.5 mA (current flowing to contact)
Allowable ON contact resistance 250 Ω max.
Allowable OFF contact resistance 100 kΩ min.
Allowable ON residual voltage
1.0 V max. (open collector type)
Allowable OFF leakage current
50 µA max. (open collector type)
Reference
voltage
output
Output rating
5.0 V DC ±5 %, 5 mA max.
Application
Reference voltage of flow rate setting voltage and 5 V input of external
3-way switching input
Communications Mode
Dedicated loader communications
RS-485 communications (3-wire system)
*7
*8
Transmission speed
2400, 4800, 9600, 19200, 38400 bps
(loader communications: 19200 bps only)
Power
Rating
24 V DC, current consumption: 400 mA max.
24 V DC current
consumption: 500 mA max.
Allowable power voltage range
(T: operating temperature)
21.6 to 26.4 V DC
(ripple 5 % max.)
(1) 21.6 to 26.4 V DC
(–10 °C≤T≤40 °C)
(2) 23.5 to 26.4 V DC
(40 °C<T≤60 °C)
(ripple 5 % max.)
(1) 21.6 to 26.4 V DC
(–10 °C≤T≤35 °C)
(2) 23.5 to 26.4 V DC
(35 °C<T≤50 °C)
(ripple 5 % max.)
Isolation
The power circuit is isolated from the input/output circuit.
Material of gas-contacting parts
SUS316, Teflon, Fluororubber
Connection method
3/4-16UNF, Rc 1/2", 1/2" Swagelok, 1/2" VCR
Mounting orientation
Horizontal. Be sure that display surface does not face down.
Mass
Approx. 3.5 kg
Applicable standards
EN61326-2-3: 2013, EN61326-1: 2013
(To be used in an industrial electromagnetic environment)
During EMC testing, the reading or output may fluctuate by the
equivalent of ±10 %FS.