43
Operating instructions XENOTEST BETA+
7.
Operation
7.4
Starting the test program
The program control of the XENOTEST BETA+ allows access to:
•
O
WN
TEST
- user specific test programs - The following test
conditions can be defined:
• phases of light or dark,
• controlled irradiation intensity,
• controlled black standard temperature,
• controlled sample chamber temperature,
• controlled humidity,
• cyclic spraying or dewing of the samples.
•
Standards - pre-programmed tests that conform to
international standards.
Selecting and starting a test program, Fig. 33:
1. After switch-on, the display
1
shows the test input
dialog window.
2. Set the configuration type
5
for the test. Specify whether
user specific or pre-programmed tests by Atlas are to be
used (Own tests / Standards).
3. Highlight the test program
4
: Use the UP / DOWN arrow
keys
2
to select the test program.
4. To start the test program:
Press the
t
key
5. To interrupt the program run:
Press the
n
key
6. To display a test overview:
Press the LEFT / RIGHT arrow keys
2
NOTE - Software documentation
The operation of the instrument control and
the programming of the test programs is
described in the separate software documen-
tation.
Abb. 33