34
Operating instructions XENOTEST BETA+
6.
Start-up
6.6
Installing the XENOSENSIV sensor
CAUTION - Risk of burns!
The test chamber is extremely heated during
operation. Contact with heated test chamber
components may cause burns. Prior to any
sensor installation or removal, allow the test
chamber to cool down sufficiently!
1.
Fig. 22:
Install the XENOSENSIV sensor
2
using the two
retaining screws
4
to the adapter
3
of the sensor mounting
bracket
1
.
2. For installation, the sensor mounting bracket
1
is positi-
oned to the left or right sample holder straps next to the
connecting sleeve
6
.
3. Connect the connector
5
of the connecting cable to the
sensor receptacle
6
.
NOTE - Operation without sensor
If the XENOTEST BETA+ is operated without
XENOSENSIV sensor, the adapter
7
for ope-
ration without sensor (terminating plug) must
be connected to the connecting sleeve
6
.
Fig. 22