37
Operating instructions XENOTEST BETA+
6.
Start-up
6.9
Setting up the interface
Fig. 25:
The instrument is equipped with four interfaces for data
communication with external systems.
Network connection
1
:
Via the protocol "Integrated Fast Ethernet Controller" (3C905C-
TX-compatible), the test instrument can be integrated into a
network. Available only with later software version.
Serial interface
2
:
The RS232 interface allows the output of measuring data to
a computer during a running test program.
USB interface
3
:
Data interface in accordance with USB 2 standard. This interface
allows the output of measuring data to a computer during a
running test program. Available only with later software
version.
Slot
4
for SmartMedia chip
5
:
For updating already installed instrument software, loading new
check programs or downloading measuring and test data.
6.10
Power supply connection
WARNING - Electric shock!
Contact with current-carrying components may
cause a lethal electric shock.
Prior to connecting the instrument to the power
supply system, check plug and power supply
cable.
Do not use damaged components for a connec-
tion to the power supply system!
The XENOTEST BETA+ is connected to a mains voltage of 400
V ± 10 %
The power supply connection is made through a grounding plug
(3P,N,PE), CEE (32 A, 5 poles, 6h).
The power supply system must be properly fused.
Connecting and switching the instrument on:
1. Prior to connecting the instrument to the power supply
system, check to see if the specifications of the power
supply system in the operating rooms corresponds with the
ratings on the nameplate at the rear of the instrument.
2. If the ratings for voltage (V) and maximum current (A) do not
correspond with the technical data (see Section 11), the
instrument must not be connected.
3.
Fig. 26:
Connect the grounding plug
2
(at the rear of the
instrument) to a properly grounded and fused outlet.
4. Make sure that the power supply cable is not subject to
pressure or tensile forces.
5.
Fig. 26:
Move the main switch
1
to position "I".
Fig. 25
Fig. 26
Содержание XENOTEST BETA+
Страница 1: ...Operating Instructions XENOTEST BETA ...
Страница 59: ...59 Operating instructions XENOTEST BETA Notes ...