10
Operating instructions XENOTEST BETA+
1.
Safety instructions
Safety devices
The XENOTEST BETA+ is equipped with safety switches and temperature sensors that
monitor the individual functions.
•
If the test chamber door is opened during operation, all functions are switched off.
•
A thermoswitch monitors the heat build-up within the test chamber.
•
A thermoswitch monitors the exhaust airflow. If the exhaust air temperature exceeds
120 °C, all functions are switched off.
•
A float switch terminates all functions whenever the water reservoir filling level falls below
a minimum quantity.
•
If the Xenon irradiator fails to ignite after four attempts, the ignition process is
cancelled.
After the instrument has been switched off due to a fault condition, the operation can be
resumed only after the failure message has been acknowledged. An interrupted test will be
resumed from the point where it had been interrupted.
Power supply connection
The XENOTEST BETA+ is supplied with electrical power through a power supply
connection with:
•
400 V ± 10 %, (3P, N, PE), CEE (32 A, 5 poles., 6 h).
Disposal
•
Please discard the packaging materials according to the applicable discarding
guidelines. For a list of the used packaging materials, please refer to Section 2.1,
"Packaging materials".
•
Used instrument contain reusable materials. Therefore, before discarding instruments,
contact your local magistrate or community office about the possibilities of recycling.
From 1 September 2004, local community discarding codes will no longer apply but
are replaced by the national implementation of the European Code
WEEE
(
W
aste
E
lec-
trical and
E
lectronic
E
quipement).
•
Xenon irradiators can be shipped to ATLAS MTT GmbH to be discarded for a fee.
All other components of the instrument can be recycled through regular channels.