27
Operating instructions XENOTEST BETA+
6.
Start-up
6.1
Initial start-up
NOTE - Initial start-up
The initial start-up must be performed only by
an ATLAS Technical Service rep or by an
authorized service company.
During the initial start-up, all functional units of
the instrument are checked for their correct
functions and operability.
The instrument is commissioned in an opera-
tional state; the commissioning is documented
in the commissioning log at the end of these
instructions.
6.2
Checking the instrument components
WARNING - Electric shock!
Contact with current-carrying components may
cause a lethal electric shock.
Before checking the instrument components,
disconnect the instrument from the power
supply system!
•
Switch the instrument off using the ON/OFF
switch.
•
Disconnect the mains plug from the socket
and protect it from accidental reconnection.
•
Check to see if the instrument is de-ener-
gized.
Only instrument components that are in perfect condition ensure the operational safety of
the XENOTEST BETA+.
Prior to any start-up or test run, check components and their functional units listed below
for possible damage and operability.
Do not use damaged or worn components.
Checklist:
The following functional units must be checked prior to any start-up.
•
Correct combination of light filters.
CAUTION - Overheating!
Apart from the light filtering, the light filters are
also used for absorbing heat. If tests are con-
ducted without light filters, the instrument may
overheat and the samples may be damaged.
Ignite the Xenon irradiators only when light
filters are employed.
•
Number of Xenon lamp operating hours.
•
Condition of the air filters for irradiator cooling and test chamber cooling.
•
Connection and possible leakage of water hoses.
•
Condition of the water filter.