Operating instructions XENOTEST BETA+
9.
Troubleshooting
Tab. 5-2
Error message
Correction
Low water level, reservoir
Top up reservoir
Sample chamber door open.
Close sample chamber door.
Device cover open.
Close device cover.
Sample chamber overtemperature.
Contact Technical Service
Irradiator exhaust air overtemperature.
Contact Technical Service
Series device overload.
Contact Technical Service
Sample chamber air recirculation faulty.
Contact Technical Service
Sprinkler failure.
Contact Technical Service
Humidification failure.
Contact Technical Service
Battery missing or discharged.
Contact Technical Service
Battery older than 5 years.
Contact Technical Service upon next inspection.
Low water level, humidifier.
Check the tubing and connections.
Check the filling level in the reservoir.
XENOSENSIV sensor or terminating
plug not installed.
Check that the XENOSENSIV sensor or terminating plug are
installed.
Check the xenon bulb seat.
Check the xenon bulb fuse.
Check the xenon bulb`s age.
Lamp will not ignite.
Contact Technical Service. The device remains operative.
Reenter program, time and filter system.
Memory error
Performing basic inizilisation
--> ENTER
All device data and calibration factors
deleted --> ENTER
Power failure
on dd.mm.yy. at hh:mm
Refer to the software documentation.
Resume interrupted program by inputting the appropriate data.