Operating instructions XENOTEST BETA+
5.
Functional description
5.4
Ventilation system
Fig. 11:
The XENOTEST BETA+ is air-cooled. The cooling system is supplied with two
different airflows: irradiator cooling
D
and test chamber cooling
6
.
Irradiator cooling:
A blower
C
aspirates the air
D
for the irradiator cooling at the air vents
3
at the right
side of the instrument. The air flows along the assembly duct and cool the electronic com-
ponents. Then, the air is routed through the external cylinder
2
of the irradiator unit and
cools the Xenon irradiators.
The heated air is carried off through the vent at the top of the device
1
and is aspirated by
a technical ventilation in or removed directly out of the laboratory.
Test chamber cooling in fresh air mode:
The air
6
for the test chamber cooling is aspirated by the blower
9
through the air filter
5
at the air inlet opening.
Depending on the present temperature in the test chamber
F
, the fresh air feed rating is
controlled using the motor-driven doors
3
and
7
.
The airflow is routed past the heating system
8
and heated to the required tempera-
ture. The humidification system
A
then mixes the air with humidified air and feeds the mixture
into the test chamber
F
. The humidification system removes air from the test chamber
through the inlet air pipe
B
.
The airflow
E
is distributed evenly around the sample turntable while temperature and
humidity are constantly measured by the integral sensor system
2
.
The used air is carried off through the exhaust air opening
4
and is aspirated by a technical
ventilation in or removed directly out of the laboratory.
Test chamber cooling in recirculating air mode:
In the recirculating air mode, the two air doors
3
and
7
are closed so that the airflow
is routed into the closed air recirculation system.
Fig. 11