1 of 1
Page: 55
Issue: 4.2.1
2SFT1-0001
Manual
3. The reading should be close to 0 Ω.
4. If not, replace the power cable.
Figure 76: Towcable TCU Connection Pinout
Figure 77: Towcable Wet-Mate Connector Pinout
5.1.3.2
Power Cable
1. Set the multimeter on
Ω
.
2. Touch each of the pins on the other end with the other probe in the following sequence.
Pin 1 -> Pin 2
Pin 2 -> Pin 1
Pin 1 -> Pin 1
Pin 2 -> Pin 2
*Pin 3 is not used.
3. The reading should be close to 0 .
4. If not, contact Marine Sonic Technology, Ltd for repair.