10.1.3
Field “Pulse Repetition”
Figure 12: Field “Pulse Repetition”
Pulse repetition:
Input an integer and select the units (ns, µs, ms) from the dropdown
box in order to set the temporal value. The meaning of this value is
determined by selection of one of the three radio buttons described
below.
Period:
Radio button to select that the numerical input is to determine the
period of the pulses of the current sequence.
Separation:
Radio button to select that the numerical input is to determine the
separation of the pulses of the current sequence.
Repetition rate:
Radio button to select that the numerical input is to determine the
repetition rate of the pulses of the current sequence.
10.1.4
Field “Triggering Mode”
Figure 13: Field “Triggering Mode”
External:
The sequence as programmed through all of the other parameters is
enabled by clicking the “Start” button on the GUI (Control section).
However, the sequence will only start when the first rising edge into
the “External Trigger” BNC-receptacle on the back panel of the
main control unit is detected.
Internal:
The sequence as programmed through all of the other parameters is
started by clicking the “Start” button on the GUI.
Содержание LDD100-F Series
Страница 25: ...LDD100 Operating Instructions 25 72 5 1 1 Safe Operating Area LDD100 F040...
Страница 27: ...LDD100 Operating Instructions 27 72 5 2 1 Safe Operating Area LDD100 F120...
Страница 29: ...LDD100 Operating Instructions 29 72 5 3 1 Safe Operating Area LDD100 L040...
Страница 31: ...LDD100 Operating Instructions 31 72 5 3 2 Safe Operating Area LDD100 L120...
Страница 33: ...LDD100 Operating Instructions 33 72 5 4 1 Safe Operating Area LDD100 L200...
Страница 35: ...LDD100 Operating Instructions 35 72 5 5 1 Safe Operating Area LDD100 XL040...
Страница 37: ...LDD100 Operating Instructions 37 72 5 6 1 Safe Operating Area LDD100 XL120...
Страница 39: ...LDD100 Operating Instructions 39 72 5 7 1 Safe Operating Area LDD100 XL200...
Страница 41: ...LDD100 Operating Instructions 41 72 5 8 1 Safe Operating Area LDD100 XL400...
Страница 43: ...LDD100 Operating Instructions 43 72 5 9 1 Safe Operating Area LDD100 XL600...