LDD100
Operating Instructions
68 / 72
If Byte 33 “!” is returned, then the uploaded parameter set is not valid.
@e – start a sequence with external Trigger. When the trigger signal rises, the full sequence is
generated.
If Byte 33 “!” is returned, then the uploaded parameter set is not valid.
@E - start a single pulse of a sequence. When the trigger signal rises only one pulse of the
sequence is generated. To complete the sequence a signal for every pulse must be applied.
NOTICE!
Use of this function requires special care in order to not damage the
instrument!
Setting the @E command at a repetition rate higher than allowed by the safe
operating area can damage the instrument. See 5.1.1 ”Safe Operating Area
(LDD100-F040)” pg. 25 and 5.2.1 ”Safe Operating Area (LDD100-F120)”
pg. 27 for details.
11.2.9
Stop a sequence
When sending an escape (Byte 27) the driver stopps immediately. It quits by sending two
bytes. The first one is the current modulator flag. The second is the internal measured
temperature of the laser driver in °C.
Byte 1 – Bit 0 to Bit 7:
Bit 2: Driver is ready
Bit 3: Excess temperature
11.2.10
Read status and temperature of the laser driver
By sending @R to the driver two bytes will be returned. These indicate the current modulator
flags (first byte) and the internal measured temperature (second byte).
See: Stop a sequence.
Содержание LDD100-F Series
Страница 25: ...LDD100 Operating Instructions 25 72 5 1 1 Safe Operating Area LDD100 F040...
Страница 27: ...LDD100 Operating Instructions 27 72 5 2 1 Safe Operating Area LDD100 F120...
Страница 29: ...LDD100 Operating Instructions 29 72 5 3 1 Safe Operating Area LDD100 L040...
Страница 31: ...LDD100 Operating Instructions 31 72 5 3 2 Safe Operating Area LDD100 L120...
Страница 33: ...LDD100 Operating Instructions 33 72 5 4 1 Safe Operating Area LDD100 L200...
Страница 35: ...LDD100 Operating Instructions 35 72 5 5 1 Safe Operating Area LDD100 XL040...
Страница 37: ...LDD100 Operating Instructions 37 72 5 6 1 Safe Operating Area LDD100 XL120...
Страница 39: ...LDD100 Operating Instructions 39 72 5 7 1 Safe Operating Area LDD100 XL200...
Страница 41: ...LDD100 Operating Instructions 41 72 5 8 1 Safe Operating Area LDD100 XL400...
Страница 43: ...LDD100 Operating Instructions 43 72 5 9 1 Safe Operating Area LDD100 XL600...