LDD100
Operating Instructions
60 / 72
:
Radio button to activate the parallel sweep mode. This mode performs
a current sweep by stepping the pulse current from the minimum set
value to the maximum set value in a number of equidistant currents as
determined by the parameter “Steps”. More than one cycle may be run
automatically via the parameter “Number of Cycles”. Each current of
the sweep is pulsed “Number of Cycles” times before the next current
is set.
:
Radio button to activate the pulse mode. The pulse mode is used to
pulse the driver at constant current with a single pulse width and pulse
separation during the whole sequence.
10.1.8
Field “Monitor”
Figure 17: Field “Monitor”
Temperature:
The temperature of the end stage is measured at the end of each
sequence and indicated in this non-editable field.
Excess temperature: If the temperature of the end stage exceeds the maximum allowed
temperature, the end stage will shut down to prevent damage and the
status will be indicated by lighting this red virtual LED. Wait until the
end stage has cooled down (10 minutes) and then restart the system.
The parameter settings will need to be changed in order to prevent the
excess temperature state from reoccurring. An excess temperature
state is very unlikely unless the end stage head is in a hot ambient
environment or convection cooling of the head is prevented in some
manner.
NOTICE!
Do not place the end stage unit in a vacuum or prevent convection
cooling!
The end stage head is convection cooled and as such must be in a free flowing
air environment.
Ready:
This green virtual LED is lit when the system is functionally operable.
Содержание LDD100-F Series
Страница 25: ...LDD100 Operating Instructions 25 72 5 1 1 Safe Operating Area LDD100 F040...
Страница 27: ...LDD100 Operating Instructions 27 72 5 2 1 Safe Operating Area LDD100 F120...
Страница 29: ...LDD100 Operating Instructions 29 72 5 3 1 Safe Operating Area LDD100 L040...
Страница 31: ...LDD100 Operating Instructions 31 72 5 3 2 Safe Operating Area LDD100 L120...
Страница 33: ...LDD100 Operating Instructions 33 72 5 4 1 Safe Operating Area LDD100 L200...
Страница 35: ...LDD100 Operating Instructions 35 72 5 5 1 Safe Operating Area LDD100 XL040...
Страница 37: ...LDD100 Operating Instructions 37 72 5 6 1 Safe Operating Area LDD100 XL120...
Страница 39: ...LDD100 Operating Instructions 39 72 5 7 1 Safe Operating Area LDD100 XL200...
Страница 41: ...LDD100 Operating Instructions 41 72 5 8 1 Safe Operating Area LDD100 XL400...
Страница 43: ...LDD100 Operating Instructions 43 72 5 9 1 Safe Operating Area LDD100 XL600...