
LDD100
Operating Instructions
64 / 72
18
Continuous Mode
0 - 1
0: Not Continuously
1: Continuously
19
Pulse Width Timer
0 – 250
0 – 150µs (600ns steps)
20
Pulse Separation Timer (High Byte)
200 - 20 000
10µs – 1ms
range multiplier for
pulse separation
21
Pulse Separation Timer (Low Byte)
Following upload of the parameter set, the instrument performs an internal test of the validity
of the input values. If the parameter set is valid, the Byte 0 is returned (no errors). If any
errors have occurred, the Byte 13 is returned followed by the error code.
Table 16: Error Codes
Error Code
Meaning
10
Period: value too low (<1)
11
Period: value too high (>
10,000)
12
Pulse duration: value too low (<1)
13
Pulse duration: value too high (>
40,000)
14
Minimum current: value too high (>
4,095)
15
Maximum current: value too high (>
4,095)
16
Number of steps: value too low (<1)
17
Number of steps: value too high (>
1,000)
18
Bias: value too high (>
4,095)
19
Number of cycles: value too high (>
1,000,000,000)
20
Number of cycles: value too low (<1)
21
Mode: value too high (>
2)
22
Continuous mode: value too high (>
1)
23
Pulse
separation
timer: value too high (>
250)
24
Pulse
separation
timer: value too low (<
200)
25
Pulse
separation
timer: value too high (>
20,000)
50
The minimum current is lower than the bias current
51
The maximum current is lower than the bias current
52
Safe operating area: the duty cycle is too high
53
Safe Operating Area: the pulse duration is too high for the maximum. current
54
The minimum current is higher than the maximum current
55
The repetition period must be at least 9µs
Содержание LDD100-F Series
Страница 25: ...LDD100 Operating Instructions 25 72 5 1 1 Safe Operating Area LDD100 F040...
Страница 27: ...LDD100 Operating Instructions 27 72 5 2 1 Safe Operating Area LDD100 F120...
Страница 29: ...LDD100 Operating Instructions 29 72 5 3 1 Safe Operating Area LDD100 L040...
Страница 31: ...LDD100 Operating Instructions 31 72 5 3 2 Safe Operating Area LDD100 L120...
Страница 33: ...LDD100 Operating Instructions 33 72 5 4 1 Safe Operating Area LDD100 L200...
Страница 35: ...LDD100 Operating Instructions 35 72 5 5 1 Safe Operating Area LDD100 XL040...
Страница 37: ...LDD100 Operating Instructions 37 72 5 6 1 Safe Operating Area LDD100 XL120...
Страница 39: ...LDD100 Operating Instructions 39 72 5 7 1 Safe Operating Area LDD100 XL200...
Страница 41: ...LDD100 Operating Instructions 41 72 5 8 1 Safe Operating Area LDD100 XL400...
Страница 43: ...LDD100 Operating Instructions 43 72 5 9 1 Safe Operating Area LDD100 XL600...