Routine Care
6-2
Angel
®
cPRP System
Operator’s Manual
3. Select the source to download the new software version. Navigate the list of available sources and
software versions using the Decrement and/or Increment Buttons. To install the most recent version
of software from the available sources, do not do anything at this step.
4. Touch the “
Download software from this source
” Button.
Enter the unlock code for the selected software version (see Figure 6-1). The software unlock
code is available from the manufacturer.
Figure 6-1 Unlock Code Screen
5. Use the keypad on the Touch Screen Display to enter the unlock code.
When the correct unlock code has been entered, the “
Continue
” Button will become enabled.
Touch the “
Cancel
” Button to return to the New Software Screen without entering an
unlock code.
6. Touch the “
Continue
” Button.
The selected software version will be copied to the
Angel System’s
temporary memory and be
validated. Three successive progress bars monitor each stage of this process: checking for
updates, copying, and validating the downloaded files.
Once the new software version has been copied to the
Angel System
and validated, the Software
Validated Screen is displayed (see Figure 6-2).
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...