Processing
3-4
Angel
®
cPRP System
Operator’s Manual
Compartment of the Three-Compartment Reservoir Bag will contain platelet
poor plasma.
Note:
Before processing more than one cycle of blood or the mixture of blood
and bone marrow, agitate the Whole Blood Compartment of the reservoir bag to
mix the remaining blood, providing a more uniform collection of blood
components.
4. The
Angel System
allows a maximum total processing volume of 180 mL. If the
maximum volume has not been reached and the available volume is greater than
40 mL, the “
New Cycle
” Button will be available. Touching the “
New Cycle
”
Button will display the Start Screen (see Step 2). Otherwise, end the case by
touching the “
End Case
” Button. Touching the “
End Case
” Button will display the
End of Case Screen, which displays the total amounts of RBC, PPP, and PRP
collected during the case (see Figure 3-5).
The End of Case Screen allows the case data to be saved (refer to “
Saving
Case Data
” on page 3-6). The End of Case Screen also allows case data to be
entered (refer to “
Entering Optional Case Data Fields
” on page 3-7).
Figure 3-5 End of Case Screen
5. Remove and properly dispose of the
Angel Processing Set
in the following way:
a. Open the Centrifuge Lid and rotate the Centrifuge to a position so that the
interlock mechanism shown in Figure 2-3, item 2 does not interfere with the
Stator Arm Brake.
NOTE: If the Stator Arm Brake interferes with the interlock mechanism, the
Separation Chamber Plate cannot be unloaded properly.
b. Lift the Centrifuge Stator Arm sufficiently to lock the Centrifuge (with the Stator
Arm Brake) within the Centrifuge Well so the Centrifuge cannot move.
Note:
Close all open ports using the Male-Female Luer lock plugs, before
removing the Processing Set from the machine.
c. Remove the Variable Volume Separation Chamber by twisting it counter-
clockwise and then lifting it out of the Centrifuge Adapter.
d. Remove the Platelet Cuvette / Valve Assembly by grasping the Platelet
Cuvette / Valve Assembly release tab and lifting it upwards. To remove the
For a detailed explanation of
the End of Case Screen,
read “
End of Case Screen
”
on page 3-17.
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...