Processing
3-26
Angel
®
cPRP System
Operator’s Manual
Stop Button
The Stop Button is not required in normal use of the
Angel System
.
However, if it becomes necessary to stop the processing cycle before
it has completed, follow these steps:
Press the Stop Button mounted on the front of the
Angel System
. This
will halt the system’s Pump, Centrifuge, and Valve, and will display the Empty Screen
on the Touch Screen Display (see Figure 3-20).
Touch the “
Empty the set
” Button to empty the Variable Volume Separation Chamber.
The Variable Volume Separation Chamber must be empty in order to remove the
disposable set from the machine.
Figure 3-20 Empty Screen
“Continue processing” Button
Touching the “
Continue processing
” Button continues processing
the current cycle without emptying the Variable Volume Separation
Chamber. This button is inactive, if the Stop Button was pressed during the blood
component collection phase.
“Empty the set” Button
Touching the “
Empty the set
” Button causes (1) the Centrifuge to
stop and (2) the pump to continue until it empties any residual fluid
from the Variable Volume Separation Chamber into the Whole Blood Compartment,
ending the current cycle. (The Variable Volume Separation Chamber must be empty in
order to remove the disposable set from the machine.)
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...