Troubleshooting
Angel
®
cPRP System
Operator’s Manual
5-5
Text
Description
Response
No USB storage device found!
The
Angel System
does not
detect any connected USB
storage device.
To output tally data or a log file to
a USB storage device, refer to
the section “
Troubleshooting
the Save Process
” on page 5-6
for troubleshooting steps.
Output failed!
Output failed while attempting to
save tally data.
Refer to the section
“
Troubleshooting the Save
Process
” on page 5-6 for
troubleshooting steps.
Save failed!
Output failed while attempting to
save the case log file. Or, a
custom protocol could not
be saved.
Refer to the section
“
Troubleshooting the Save
Process
” on page 5-6 for
troubleshooting steps.
Past case data lost!
Some or all of the data for past or
current cases has been lost.
No response necessary. The past
case data may not be recovered.
Power fail protocol not found:
{Protocol Name}
A power fail has occurred, and
the given protocol cannot
be found.
The
Angel System
will
automatically select the default
protocol for use. If a different
protocol is desired, select it from
the Protocol Pull-Down Button.
The lost protocol may need to be
recreated before it can be used.
Power was lost before the end
of the case.
The
Angel System
was shut
down or lost power before the
end of a processing case.
If the power was lost in the
middle of a cycle, then the set
must be emptied. A new cycle or
case may then be started.
Redo failed! Undo/redo history
cleared.
An attempt to redo a change
while creating a protocol has
failed.
The undo/redo history has been
cleared. A redo command is not
possible at this time.
Rename failed!
An attempt to rename a protocol
has failed.
Try renaming the protocol again.
Settings lost and were set to
default!
The
Angel System’s
settings
were lost, and the default settings
are being used.
To use settings other than the
default settings, make changes
using the “
Settings
” tab of the
Menu Screen (see “
Settings
Tab
” on page 3-20).
Software update is not valid!
The new software downloaded
from the selected source is
corrupt.
Try downloading the software
again. If the problem continues,
download the software from a
different source.
The software did not start
successfully.
An error occurred while the
Angel System
was powering on.
Turn the power off and then on. If
the problem continues, contact
Arthrex, Inc. Customer Service.
This version of software is
already installed on this
machine.
The operator attempted to install
the same software that is already
installed on the machine.
Choose a different version of
software to install.
Undo failed! Undo/redo
history cleared.
An attempt to undo a change
while creating a protocol has
failed.
The undo/redo history has been
cleared. An undo command is not
possible at this time.
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...