Troubleshooting
5-4
Angel
®
cPRP System
Operator’s Manual
Text
Description
Response
Another protocol has this
name–will be replaced
The currently displayed protocol
name (in the Keyboard Screen)
conflicts with an existing protocol.
If saved, the new protocol will
replace the existing protocol.
Modify the new protocol name so
that it is unique and doesn’t
overwrite the existing protocol.
Bad protocol:
{Protocol Name}
The given protocol is corrupt and
cannot be used.
The
Angel System
will
automatically select the default
protocol for use. To use a
different protocol, select it from
the Protocol Pull-Down Button.
The protocol with the error will
have to be recreated before it can
be used.
Copy failed!
An attempt to copy a protocol
has failed.
Try copying the protocol again.
Corrected invalid protocol:
{Protocol Name}
The given protocol was invalid.
No response is needed. The
Angel System
automatically
corrected the protocol.
Delete failed!
An attempt to delete a protocol
has failed.
Try deleting the protocol again.
Download failed!
The download of a software
update has failed.
Try downloading the software
again. If the failure continues,
use a different download source
or contact Arthrex, Inc.
Customer Service.
Hardware does not support
this version of software!
The operator tried to install a
software version that is not
supported by the current
hardware.
Select a previous version of
software that is compatible with
the current hardware.
Hardware malfunction!
A hardware malfunction
has occurred.
Cycle the machine power off
and then on. If the problem
continues, contact Arthrex, Inc.
Customer Service
Installation failed!
Installation failed while
downloading new software.
See “
Troubleshooting the
Software Update Process
” on
page 5-6 for troubleshooting
steps.
Insufficient memory to load
this version of software!
The operator tried to install a
software version that is not
supported by the current
hardware.
Select a previous version of
software that is compatible with
the current hardware.
No destination found!
The
Angel System
is not
connected to a valid destination
or the destination cannot
be reached.
To output a log file to an external
destination, refer to the section
“
Troubleshooting the Save
Process
” on page 5-6 for
troubleshooting steps.
No software updates found!
No software updates were found. See “
Troubleshooting the
Software Update Process
” on
page 5-6 for troubleshooting
steps.
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...