Processing
3-6
Angel
®
cPRP System
Operator’s Manual
Saving Case Data
The
Angel System
stores a tally table and a detailed log file about every case that has
been processed, with an option to save the tally table and the detailed case log file.
The tally tables and case logs may be saved to USB storage devices. The case logs
may be saved to a USB storage device.
Note:
Only devices or cables meeting IEC 60950 and IEC 60601-1 should be
connected to the USB Port and/or Ethernet Port(if present). Failure to do so may result
in operator shock. All cables used in conjunction with the device should be no longer
than 1 m (3 ft.) in length. Operators connecting other devices to the USB Port must
ensure compliance to the system requirements of IEC 60601-1.
Every case includes the following information, which is available as an easily
viewable and printable text file:
•
The case number, a sequential number assigned by the machine to every case;
the
Angel System
serial number; the date and time at which the case was started;
and the
Angel System
software version number.
•
The “
ID
” field, as entered by the operator (refer to “
Entering Optional Case Data
Fields
” on page 3-7).
•
The “
Operator
” field, as entered by the operator (refer to “
Entering Optional Case
Data Fields
” on page 3-7).
•
The tally table: the volumes (mL) of the whole blood or the blood and bone marrow
mixture processed by the case, as well as volumes of RBC, PPP, and PRP
collected during each cycle of the case. When available, the start and end times of
the case, as well as the protocol used during processing are included.
•
The “
Notes
” field, as entered by the operator (refer to “
Entering Optional Case
Data Fields
” on page 3-7).
The optional fields may be left blank. The fields are displayed as blank lines.
Once the processing of a case has been completed, there are two places within the
Touch Screen user interface where the case data may be saved:
•
The End of Case Screen, which appears upon completion of each case
(see Figure 3-5 above).
•
The “
Tally
” tab of the Menu Screen (see Figure 3-6).
Содержание Angel
Страница 2: ...This page intentionally left blank ...
Страница 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 14: ...Overview xii Angel cPRP System Operator s Manual ...
Страница 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Страница 18: ......
Страница 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 76: ...Programmability Option 4 10 Angel cPRP System Operator s Manual This page intentionally left blank ...
Страница 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...