5.2.1 SillHa Software [Warning] Causes and Remedies
A warning occurs as a result of an incorrect operating procedure. You can
continue using the instrument by following simple remedial procedures.
No.
Description
Cause
Remedy
[W-301]
The
corresponding
patient
information
does not exist.
Data corresponding
to the entered
medical chart number
or patient name does
not exist.
Click [OK] and enter the correct medical
chart number or patient name.
[W-302]
Enter the
medical chart
number.
The medical chart
number has not been
entered.
Click [OK] and enter the medical chart
number.
[W-303]
The
corresponding
medical chart
number already
exists.
The same medical
chart number has
already been
registered.
Click [OK] and confirm the correct
medical chart number.
[W-304]
The
corresponding
test history
does not exist.
Data corresponding
to the entered
medical chart number
does not exist.
Click [OK] and enter the correct medical
chart number.
[W-305]
Specify the
export-to
location.
The export-to location
of the data has not
been specified.
Click [OK] and specify the export-to
location.
[W-306]
Specify the
import-from
location.
The location where
the import-from data
is saved has not been
specified.
Click [OK] and specify the storage
location of the data.
[W-307]
The password
is incorrect.
The password is
incorrect.
Click [OK] and enter the correct
password.
[W-308]
Select the data. [Edit] was clicked
even when the result
sheet was not called.
Call the result sheet to be edited, and
click [Edit].
[W-309]
Select up to
two records.
Three or more past
test results have
been selected for
displaying in the
result sheet (Type A
or B).
Click [OK]. Select up to two past test
results.
[W-310]
No
measurement
data was
found.
The necessary
measurement data
could not be
received.
Click [OK]. Tap [History] on the
instrument screen, make sure the
measurement results are for the
corresponding medical chart number,
and tap [Transfer]. If the measurement
results are not for the corresponding
medical chart number, or if the results
cannot be transferred, retry
measurement.
[W-311]
The target data
does not exist.
The measurement
results to be used as
transmission data
have not been saved.
Click [OK], perform measurement, and
create the transmission data again.