SillHa Oral Wellness Instrument
5.2.2 SillHa Software [Error] Causes and Remedies
An error occurs due to incorrect operations performed as a result of faulty PC
environment. You can continue using the instrument by removing the cause of the
error.
No.
Description
Cause
Remedy
[E-401]
The update
has failed.
An error occurred
while the database
was being updated.
Click [OK] and again perform the
operation that was in progress before
the occurrence of the error.
[E-402]
The
transmission
has failed.
An error occurred
during
communication.
Reconnect the USB cable between
the PC and the instrument. Restart the
SillHa software and the instrument,
and repeat the operation you were
performing before the error occurred.
[E-403]
Data could
not be
exported.
The PC memory is
insufficient.
Click [OK]. Expand the PC memory.
[E-404]
Data could
not be
imported.
Either data does not
exist in the
specified import
from location, or the
file is corrupted.
Click [OK] and check the storage
location of the data to be imported.
[E-405]
The test
results list
data could
not be
created.
The storage
capacity (HDD) of
the PC is
insufficient.
Click [OK]. Expand the PC memory.
[E-406]
Printing could
not be
performed.
The printer is not
connected to the
PC.
Properly connect the printer to the PC.
[E-407]
The
measurement
data could
not be
received.
Data could not be
received properly
from the instrument
at the end of
measurement
Click [OK]. Properly connect the
instrument to the PC. Tap [History] on
the instrument, make sure the
measurement results are for the
corresponding medical chart number,
and tap [Transfer]. If the measurement
results are not for the corresponding
medical chart number, or if the results
cannot be transferred, retry
measurement.