30
ARKRAY
Place the test strip on the test strip holder.
• Slide the test strip along the groove in
the test strip holder until it stops.
Close the reading unit cover.
REFERENCE
If the test strip has not been placed until the
test strip holder starts moving: [E-100]
occurs. Tap [OK] and repeat the procedure
from step 1 on page 27, “2.3.2. Enter the
Patient Information on the PC”.
Measurement will start automatically, and the remaining time is displayed.
IMPORTANT:
If the test strip deviates from the groove of
the test strip holder, accurate measurement
results cannot be obtained.
The test strip holder will start moving. Do not touch the instrument
until measurement is complete to prevent personal injury. The test
strip holder moves twice, when the remaining time is approximately
260 sec. and approximately 20 seconds.
Test strip
Reading
unit cover
Reading
unit cover