Electromagnetic Flowmeter M-910
36
Operation manual
Impulse output mode setting
SPM(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 }
Following modes can be set:
•
0 Off
•
1 Pos.Flow
•
2 Neg.Flow
•
3 Bip.Flow
•
4 On Pos.
•
5 On Neg.
•
6 On In
•
7 On Out
•
8 Dose On
•
9 Dose Off
•
10 On>F1
•
11 On<F1
M910 confirms execution with string „Ok<cr>”.
Example RS232:
Command „SPM1<cr>” sets mode “Positive flowrate” for impulse output. If query „SPM?<cr>” is sent, flowmeter
returns response in format „1<cr>”.
Status output mode setting
SSM(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 }
Following modes can be set:
•
0 Off
•
1 On Pos.
•
2 On Neg.
•
3 On In
•
4 On Out
•
5 Dose On
•
6 Dose Off
•
7 On>F1
•
8 On<F1
M910 confirms execution with string „Ok<cr>”.
Example:
Command „SSM1<cr>” sets mode “On for positive flowrate” for status output. If query „SSM?<cr>” is sent,
flowmeter returns response in format „1<cr>”.
Digital input mode setting
SIM(?)<CPD> { 0 | 1 | 2 }
Following modes can be set:
•
0 Off
•
1 Dose
•
2 Clr.Vol
M910 confirms execution with string „Ok<cr>”.
Example: