Electromagnetic Flowmeter M-910
30
Operation manual
Following values are pre-set:
Datalogger standard factory setting:
Datalogger
OFF
5.2.3 General settings (3 GENERAL)
For general settings configuration or for reading actual settings. After pushing “UP” key next item is selected.
After pushing “ENTER” key following submenu is displayed:
5.2.3.1
Diameter (3.1 DIAMETER)
Flowmeters nominal diameter is displayed. After pushing “UP” key “Range” is displayed.
5.2.3.2
Nominal flowrate range Q
N
(3.2 RANGE)
Nominal flowrate range Q
N
is displayed in flowrate units. After pushing “UP” key “Serial number” is displayed.
5.2.3.3
Serial number (3.3 SERIAL NR.)
Flowmeters Serial number is displayed. After pushing “UP” key “Power supply” is displayed.
5.2.3.4
Power supply (3.4 POWER SUP.)
Information about power supply (voltage and frequency) is displayed. After pushing “UP” key “Self test” is
displayed.
5.2.3.5
Self-test (3.5 SELFTEST)
Function enables you to switch an internal self-test (flowrate simulator) “On” or “Off” after pressing ENTER key.
With “Up” key any item from the list “On” and, “Off” can be set. To change the current valid self-test state press
the ENTER key. ESC discards changes.
Note: Self-test “Off” state is normal working state of flowmeter. After switching self-test to “On” state, internal
flowrate simulator is inserted instead of the pipe. Function can be used for signal converter testing.
Flowrate in range …. Is displayed, if signal converter is OK.
Following values are pre-set:
Self-test standard factory setting:
Self-test
Off
5.2.3.6
Current Loop Test (3.6 C.LOOP TEST)
Function enables you to switch an internal test of the connected current loop “On” or “Off” after pressing the
ENTER key. With the “Up” key any item from the list “On” and, “Off” can be set. To change the current valid
Current Loop Test state press the ENTER key. ESC key discards changes.
Note: If Current Loop Test is in “On” state and current output flows less than 3 mA, error message “01 –
Current Output” will be displayed.
Following values are pre-set: