Electromagnetic Flowmeter M-910
32
Operation manual
nominal value to the new nominal value press the ENTER key. After entering the nominal value a new calibration
constant can be set. ESC key discards changes.
Following values are pre-set:
Calibration point 1 standard factory setting:
Nominal point
10% of required Q
N
Cal. Constant
is assigned according to the calibration
5.2.4.3
Calibration point 2 (4.3 CAL.POINT 2)
Function enables you to change the nominal and calibration value of Calibration point 2 after pressing the ENTER
key. With “Up” and “Right” key any value in the range of real flowrates can be set. To change the current valid
nominal value to the new nominal value press the ENTER key. After entering the nominal value new calibration
constant can be set. ESC key discards changes.
Following values are pre-set:
Calibration point 2 standard factory setting:
Nominal point
70% of required Q
N
Cal. Constant
is assigned according to the calibration
5.2.4.4
Calibration point 3 (4.4 CAL.POINT 3)
Function is available only if 3 or 4 calibration points are selected. Function enables you to change the nominal and
calibration value of Calibration point 3 after pressing the ENTER key. With the “Up” and “Right” key any value in
the range of real flowrates can be set. To change the current valid nominal value to the new nominal value press the
ENTER key. After entering the nominal value a new calibration constant can be set. ESC key discards changes.
Following values are pre-set:
Calibration point 3 standard factory setting:
Not used.
5.2.4.5
Calibration point 4 (4.5 CAL.POINT 4)
Function is available only if 4 calibration points are selected. Function enables you to change the nominal and
calibration value of Calibration point 4 after pressing the ENTER key. With the “Up” and “Right” keys any value
in the range of real flowrates can be set. To change the current valid nominal value to the new nominal value press
the ENTER key. After entering the nominal value a new calibration constant can be set. ESC key discards changes.
Following values are pre-set:
Calibration point 4 standard factory setting:
Not used.
5.2.4.6
Calibration Password (4.6 PASSWORD CA.)
Function enables you to enter a new password for calibration menu access after pressing the ENTER key. With the
“Up” and “Right”keys any password in the range between 00000 and 99999 can be set. To change the current valid
password to the new password press the ENTER key. ESC key discards changes.