Indications
TheraKair Visio
MRS is indicated for:
•
Edema management
•
Prevention or treatment of up to stage IV pressure ulcers
•
Patients requiring a surface to assist pain management
Contraindications
Patient conditions for which the application of pressure relief therapy on
TheraKair Visio
MRS is
contraindicated include:
•
Spinal cord injury
•
Cervical traction
Intended Care Setting
Intended care settings of the
TheraKair Visio
MRS are:
•
Acute Care
•
Post Acute Care
Connecting the System to Other Devices
Devices That May Be Attached to This System
None.
Devices to Which This System May Be Attached
The
TheraKair Visio
056FDQEH¿WWHGWRPRVWVWDQGDUGKRVSLWDOEHGV
The
TheraKair Visio
7KHUDS\8QLWFDQEH¿WWHGWRWKHIRRWERDUGRIPRVWKRVSLWDOEHGV
Risks and Precautions
Mattress Replacement
The
TheraKair Visio
056LVDPDWWUHVVUHSODFHPHQWV\VWHPDQGVKRXOGQHYHUEH¿WWHGEHIRUH
¿UVWUHPRYLQJWKHH[LVWLQJPDWWUHVVIURPWKHEHG)LWWLQJWKH
TheraKair Visio
mattress on top of an
existing mattress will raise the patient above the side rails of the bed, placing them at risk of falling
out of the bed or becoming trapped between the mattress and the side rails.
Transfer
Standard precautions should be taken during patient transfer.
Side Rails and Restraints
WARNING:
Use or non-use of restraints, including side rails, can be critical to patient safety.
Serious or fatal injury can result from the use (potential entrapment) or non-use (potential patient
falls) of side rails or other restraints. See related
Safety Information
.