29
Guidance and manufacturer’s declaration - electromagnetic immunity
The
AtmosAir
MRS
is intended for use in the electromagnetic environment specified below. The customer or user of
the
AtmosAir
MRS should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
In accordance with IEC
60601-1-2:2007, floors
are covered with synthetic
material, the relative
humidity should be at least
30%
Electrical fast
transient / burst
IEC 61000-4-4
120VAC Pump
(US)
230VAC Pump
(UKI)
±1 kV
±2 kV
±1kV
±2 kV
N/A
Surge
IEC 61000-4-5
1 kV line(s) to Line(s)
2 kV line(s) to Earth
1 kV line(s) to Line(s)
2 kV line(s) to Earth
N/A
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
5% half cycle
40% for 5 cycle
70% for 25 cycle
5% for 5 seconds
5% half cycle
40% 5 cycles
70% 25 cycles
5% for 5 seconds
N/A
Power frequency
(50Hz/60Hz)magnetic field
IEC 61000-4-8
3 A/M
3 A/M
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U, is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications
equipment and the
AtmosAir
MRS
The
AtmosAir
MRS is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the
AtmosAir
MRS can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
AtmosAir MRS
as recommended below, according to the maximum output power of the communications equipment.
Rated
maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
meters
150 kHz to 80
MHz
not applicable
80 MHz to 800
MHz
d=[ — ]
800 MHz to 2.5 GHz
d=[ — ]
0.01
N/A
0.12
0.23
0.1
N/A
0.37
0.74
1
N/A
1.2
2.3
10
N/A
3.7
7.4
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separate distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1, At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from surfaces, objects and people.
3.5
E
1
P
7
E
1
P
Содержание AtmosAir SAT 4000 Series
Страница 4: ...iv Intentionally left blank...
Страница 6: ...2 Intentionally left blank...
Страница 37: ...33 Intentionally left blank...
Страница 38: ...34 Intentionally left blank...