17
Troubleshooting
Do not attempt troubleshooting outside this guide or where the solution recommends to
contact Arjo. Any unauthorized service, modification, alteration or misuse may lead to
serious injury and / or product damage and will void all applicable warranties.
SYMPTOM
POSSIBLE CAUSE
SOLUTION
Green lamp on pump does not light
when power switch is set to the ON
position.
Power cord may be unplugged.
Plug in power cord.
Wall outlet may be controlled by a
wall switch.
Plug power cord into different wall
outlet.
Circuit breaker for wall switch may
be tripped or blown.
Reset circuit breaker or replace fuse.
Power cord or power switch may be
damaged.
Contact Arjo for assistance.
Mattress inflates only to one side
when in rotation mode.
Tubing not connected properly.
Ensure red hoses are connected
to hoses marked with red tag for
rotation.
Tubing kinked.
Check tubing inside mattress for
kinks.
Tubing disconnected.
Check tubing inside mattress for
possible disconnect.
Hole in rotation bladder.
Check deflated rotation bladder for
holes.
Mattress too firm upon arrival.
Difference in altitude not sufficient to
open valves.
Apply weight to mattress to open
valves.
Mattress cover too wrinkled upon
removal from shipping container.
Internal components have not
accommodated to environment. This
does not affect inflation or function.
Let mattress accommodate for 24
hours. If problem continues, contact
Arjo for assistance.
Mattress is not firm.
Tubing not connected properly.
Check tubing inside mattress for
loose connectors.
Tubing kinked.
Check tubing inside mattress for
possible kinks.
Tubing disconnected.
Check tubing inside mattress for
possible disconnect.
Holes in or damaged to
SAT
system.
Check
SAT
system for holes
or damage, or contact Arjo for
assistance.