background image

 5

Side Rails / Patient Restraints

 - Whether and how to use side rails or restraints is 

a decision that should be based on each patient’s needs and should be made by the 

patient and the patient’s family, physician and caregivers, with facility protocols in 

mind. Caregivers should assess risks and benefits of side rail / restraint use (including 

entrapment and patient falls from bed) in conjunction with individual patient needs, and 

should discuss use or non-use with patient and / or family. Consider not only the clinical 

and other needs of the patient but also the risks of fatal or serious injury from falling out of 

bed and from patient entrapment in or around the side rails, restraints or other accessories. 

In the US, for a description of entrapment hazards, vulnerable patient profile and guidance 

to further reduce entrapment risks, refer to FDA’s Hospital Bed System Dimensional 

and Assessment Guidance To Reduce Entrapment. Outside the US, consult the local 

Competent Authority or Government Agency for Medical Device Safety for specific local 

guidance. Consult a caregiver and carefully consider the use of bolsters, positioning aids 

or floor pads, especially with confused, restless or agitated patients. It is recommended 

that side rails (if used) be locked in the full upright position when the patient is unattended. 

Make sure a capable patient knows how to get out of bed safely (and, if necessary, how 

to release the side rails) in case of fire or other emergency. Monitor patients frequently to 

guard against patient entrapment.

CAUTION: When selecting a standard mattress, ensure the distance 

between top of side rails (if used) and top of mattress (without 

compression) is at least 8.66 in (220 mm) to help prevent inadvertent 

bed exit or falls. Consider individual patient size, position (relative to 

the top of the side rail) and patient condition in assessing fall risk. 

I.V. and Drainage Tubes –

 I.V. and drainage tubes should always have slack for 

alternating pressure or rotation and other patient movements.

Skin Care –

 Monitor skin conditions regularly and consider adjunct or alternative therapies 

for high acuity patients. Give extra attention to any possible pressure points and locations 

where moisture or incontinence may occur or collect. Early intervention may be essential 

to preventing skin breakdown.

Fluids –

 Avoid spilling fluids on pump controls. If spills do occur, clean fluid from pump 

wearing rubber gloves or while unit is unplugged to avoid any possibility of shock. Once 

fluid is removed, check operation of components in area of spill.

Fluids remaining on controls can cause corrosion, which may cause 

components to fail or operate erratically, possibly producing potential 

hazards for patient and staff.

Avoid Fire Hazards –

 To minimize risk of fire, connect the bed’s power cord directly into a 

wall-mounted outlet. Do not use extension cords or multiple outlet strips. In the US, review 

and follow FDA’s Safety Tips for Preventing Hospital Bed Fires.

No Smoking in Bed –

 Smoking in bed can be dangerous. To avoid the risk of fire, 

smoking in bed should never be allowed.

Power Cord –

 Position power cord to avoid a tripping hazard and / or damage to the cord. 

Ensure power cord is kept free from all pinch points and moving parts and is not trapped 

under casters. Improper handling of the power cord can cause damage to the cord, which 

may possibly produce risk of fire or electric shock. 

General Protocols –

 Follow all applicable safety rules and institution protocols concerning 

patient and caregiver safety.

Содержание AtmosAir SAT 4000 Series

Страница 1: ...s AtmosAit with SAT 9000 series AtmosAir with SAT APod 25 series AtmosAir with SAT T series AtmosAir with SAT V series AtmosAir with SAT M series AtmosAir INSTRUCTIONS FOR USE 407384 AH Rev 05 09 2019...

Страница 2: ...or its affiliates As our policy is one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part...

Страница 3: ...ted warranty included in this publication or with this product Information in this publication may be subject to change at any time Contact Arjo for updates Important Information For Users In order fo...

Страница 4: ...iv Intentionally left blank...

Страница 5: ...Skin Care 12 Incontinence Drainage 12 General Operation 12 Care and Cleaning 13 Fire Barrier 13 Cover Cleaning Options 13 Cleaning AtmosAir Pump and Tubes A and AR models 14 Preventive Maintenance Sc...

Страница 6: ...2 Intentionally left blank...

Страница 7: ...ating pressure and rotation AtmosAir with SAT V Series non powered AtmosAir with SAT M Series non powered The AtmosAir MRS comes in a wide variety of lengths and widths to support various customer req...

Страница 8: ...ry how to release the side rails in case of fire or other emergency Turning CAUTION Prior to engaging turn feature ensure that bed frame has side rails and that all side rails are in their full up and...

Страница 9: ...ut compression is at least 8 66 in 220 mm to help prevent inadvertent bed exit or falls Consider individual patient size position relative to the top of the side rail and patient condition in assessin...

Страница 10: ...c should be sorted as combustible waste Mattresses at the end of life should be disposed of as waste according to the national or local requirements which may be landfill or combustion Pump units have...

Страница 11: ...tress surface for staining and soiling clean and or disinfect as required see Care and Cleaning 5 Level bed and lock brakes 6 Remove existing mattress from bed frame Mattress Installation AtmosAir wit...

Страница 12: ...head end of bed frame as shown below Figures 2 and 3 4 Slip other end of locking bar into other slot on head end of bed frame as shown below Figure 4 5 Figure 3 Figure 2 Locking Bar Bed Frame Locking...

Страница 13: ...re the AtmosAir with SAT M Series MRS to the bed frame Figure 6 3 Ensure there are no gaps between mattress and bed frame or side rails Knee Bar Buckled Straps Figure 6 CAUTION When using the cardiac...

Страница 14: ...otation Attach blue hoses to the mattress connectors on the patient left side marked with BLUE tag of the mattress for alternating pressure Attach red hoses to the mattress connectors on the patient r...

Страница 15: ...rs 1 Green LED is lit when the pressure is at the set level 2 Yellow LED is lit when pressure is below the set level Check connections AtmosAir MIN POWER LOW PRESSURE NORMAL PRESSURE OFF ON MAX ALT PR...

Страница 16: ...tation The level of alternating pressure or rotation can be adjusted for patient comfort Prior to engaging alternating pressure or rotation ensure that bed frame has side rails and that all side rails...

Страница 17: ...ransport there are two options 1 Disconnect the tube assembly from the mattress Valves will close off as soon as the tube assembly is disconnected 2 Disconnect the tube assembly from the pump and secu...

Страница 18: ...ss with the fire barrier exposed to avoid contact with loose particles To ensure optimal product performance avoid over handling of the fire barrier and handle with care Fire barrier should be replace...

Страница 19: ...ls 1 Ensure pump is unplugged 2 Wipe surface of pump and tubes with coarse cloth using an approved disinfectant germicide mixed to manufacturer s instructions Do not flood any part of the pump with cl...

Страница 20: ...and hoses of A and AR models should be cleaned weekly Inspection System Check Out Check each of the following before placing the AtmosAir MRS with a new patient 1 Check mattress surface for tears or...

Страница 21: ...ed hoses are connected to hoses marked with red tag for rotation Tubing kinked Check tubing inside mattress for kinks Tubing disconnected Check tubing inside mattress for possible disconnect Hole in r...

Страница 22: ...rotation bladders for A and AR models are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems RF Welded Mattress Seperate Co...

Страница 23: ...and AR models are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems Foam SAT Cover part of SAT system SAT Interliner part...

Страница 24: ...dels are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems Foam SAT Interliner part of SAT system image for reference only...

Страница 25: ...t fits over the foam and SAT assemblies The SAT Cover SAT Cells SAT Interliner are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of S...

Страница 26: ...22 Parts Diagram AtmosAir Pump A and AR models See Replacement Parts for a complete list of parts that may be ordered for the AtmosAir Pump as well as entire pump systems...

Страница 27: ...00 9000 Foam Shell KA4 9FS SIZE APod 25 Foam Shell 25PFS SIZE T Series Foam Shell KA9FSTC V Series Foam Shell KVFS3584 M Series Foam Shell 416119 M Series Orthopedic Foam Shell 416397 AtmosAir with SA...

Страница 28: ...2 Top Vinyl Gray Bottom A9CRLVG WWWLL TXS Complete Cover Barrier Flex 1 Top Non Skid Bottom A9CBFNB WWWLL PXS Complete Cover Barrier Flex 1 Top Non Skid Bottom No Handles A9CBFNH WWWLL PXS Complete Co...

Страница 29: ...n Skid PU Bottom With Handles ADCPRPU WWWLL NXS Complete Cover Soflux Top Non Skid Bottom ADCSONB WWWLL NXS Complete Cover Soflux Top Vinyl Gray Bottom ADCSOVG WWWLL NXS Complete Cover Reliant IS2 Top...

Страница 30: ...ernating Pressure Complete Cover Reliant IS2 top and Non Skid base TXCRLNB34079TXS AtmosAir V Series PART TYPE CATALOG NUMBER V Series Mattress Complete Cover Soflux top and Vinyl Gray base AVCSOVG350...

Страница 31: ...Length 75 in 190 5 cm Maximum Length 84 in 213 cm Minimum Width 32 in 81 5 cm Maximum Width 38 in 96 5 cm Height 7 in 17 5 cm Pump 120V Weight 5 lb 2 5 kg Dimensions 10 25 in 26 cm L x 4 5 in 11 5 cm...

Страница 32: ...to the EMC information in the following tables The following tables document compliance levels and guidance from the IEC 60601 1 2 2007 Standard for the electromagnetic environment in which the Atmos...

Страница 33: ...rior to application of the test level Recommended separation distances between portable and mobile RF communications equipment and the AtmosAir MRS The AtmosAir MRS is intended for use in an electroma...

Страница 34: ...liance level in each frequency range see note b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOT...

Страница 35: ...Foot End Possible electrical shock hazard Warning of possible hazard to system patient or staff Attention See User Guide Consult User Guide On Power Off Manufacturer CE marking indicating conformity...

Страница 36: ...oxygen or nitrous oxide Serial Number The operator must read this document this instruction for use before use Note the symbol is blue on the product label Customer Contact Information For questions r...

Страница 37: ...33 Intentionally left blank...

Страница 38: ...34 Intentionally left blank...

Страница 39: ...mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Office 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 48...

Страница 40: ...ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to conti...

Отзывы: