15
3.
Wipe and rinse any soiling from the mattress surface and base. Use 1000 ppm
chlorine or 70% alcohol.
4.
Rinse with a wet towel to remove chemicals.
5.
After wipe down, dry surface with towel.
6.
Ensure bed linens are refitted and not wrinkled under patient.
7.
Clean pump and tubes (if needed) by wiping with a damp cloth.
Laundering Detachable Top Cover
1.
Unzip the top cover from the base for laundering. Do not launder base as damage
may occur.
2.
Recommended wash temperature for top cover is 60° C (140° F) for 15 minutes.
3.
Maximum wash temperature is 95° C (203° F) for 15 minutes.
4.
Tumble dry at 60° C (140° F) or air dry.
5.
Maximum drying temperature 80° C (176° F).
Cleaning
AtmosAir
Pump and Tubes (A and AR models)
1.
Ensure pump is unplugged.
2.
Wipe surface of pump and tubes with coarse cloth using an approved disinfectant
germicide mixed to manufacturer’s instructions.
Do not flood any part of the
pump with cleaning solution.
Avoid spilling fluids on pump controls. If spills do occur, clean fluid
from pump wearing rubber gloves or while unit is unplugged to avoid
any possibility of shock. Once fluid is removed, check operation of
components in area of spill. Fluids remaining on controls can cause
corrosion, which may cause components to fail or operate erratically,
possibly producing potential hazards for patient and staff.
3.
Wipe off excess solution.
4.
Disinfect pump and tubes with a chlorine solution (mixed according to the
instructions in the
Care and Cleaning
section). Using a clean cloth, wring out
excess solution until cloth is damp. Wipe pump surface and tubes with damp
cloth.
5.
Allow to air dry.