Preventive Maintenance Schedule
Preventive maintenance for the
AtmosAir
MRS consists of regular cleaning (see
) and an overall system check-out to be performed at the intervals described
below.
All components must be cleaned, disinfected and inspected after each patient’s use
and before use by a new patient.
Always use standard precautions, treating all used
equipment as contaminated. Institutions should follow local protocols for cleaning and
disinfection.
Daily Cleaning
The cover should be wiped daily with a mild soap and water solution. After wipe down, dry
surface with a towel.
Weekly Cleaning
The pump and hoses of A and AR models should be cleaned weekly.
Inspection / System Check-Out
Check each of the following before placing the
AtmosAir
MRS with a new patient:
1.
Check mattress surface for tears or cracking; do not use if tears or cracks are
present.
2.
Ensure mattress is free of stains and is not overly faded.
For A and AR models:
1.
Ensure air inlet hoses and connectors on mattress and pump are clean and
undamaged.
2.
Ensure pump and power cord are clean and undamaged.
3.
Ensure pump hanger brackets are secure and operate correctly.
4.
Ensure power switch and comfort control knob both operate correctly.
5.
Attach pump to the red
rotation
hoses and power on to ensure that the mattress
surface tilts and there are no air leaks.
6.
Attach pump to the blue
alternating pressure
hoses and power on to ensure
there are no air leaks.