background image

34 

Intentionally left blank

Содержание AtmosAir SAT 4000 Series

Страница 1: ...s AtmosAit with SAT 9000 series AtmosAir with SAT APod 25 series AtmosAir with SAT T series AtmosAir with SAT V series AtmosAir with SAT M series AtmosAir INSTRUCTIONS FOR USE 407384 AH Rev 05 09 2019...

Страница 2: ...or its affiliates As our policy is one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part...

Страница 3: ...ted warranty included in this publication or with this product Information in this publication may be subject to change at any time Contact Arjo for updates Important Information For Users In order fo...

Страница 4: ...iv Intentionally left blank...

Страница 5: ...Skin Care 12 Incontinence Drainage 12 General Operation 12 Care and Cleaning 13 Fire Barrier 13 Cover Cleaning Options 13 Cleaning AtmosAir Pump and Tubes A and AR models 14 Preventive Maintenance Sc...

Страница 6: ...2 Intentionally left blank...

Страница 7: ...ating pressure and rotation AtmosAir with SAT V Series non powered AtmosAir with SAT M Series non powered The AtmosAir MRS comes in a wide variety of lengths and widths to support various customer req...

Страница 8: ...ry how to release the side rails in case of fire or other emergency Turning CAUTION Prior to engaging turn feature ensure that bed frame has side rails and that all side rails are in their full up and...

Страница 9: ...ut compression is at least 8 66 in 220 mm to help prevent inadvertent bed exit or falls Consider individual patient size position relative to the top of the side rail and patient condition in assessin...

Страница 10: ...c should be sorted as combustible waste Mattresses at the end of life should be disposed of as waste according to the national or local requirements which may be landfill or combustion Pump units have...

Страница 11: ...tress surface for staining and soiling clean and or disinfect as required see Care and Cleaning 5 Level bed and lock brakes 6 Remove existing mattress from bed frame Mattress Installation AtmosAir wit...

Страница 12: ...head end of bed frame as shown below Figures 2 and 3 4 Slip other end of locking bar into other slot on head end of bed frame as shown below Figure 4 5 Figure 3 Figure 2 Locking Bar Bed Frame Locking...

Страница 13: ...re the AtmosAir with SAT M Series MRS to the bed frame Figure 6 3 Ensure there are no gaps between mattress and bed frame or side rails Knee Bar Buckled Straps Figure 6 CAUTION When using the cardiac...

Страница 14: ...otation Attach blue hoses to the mattress connectors on the patient left side marked with BLUE tag of the mattress for alternating pressure Attach red hoses to the mattress connectors on the patient r...

Страница 15: ...rs 1 Green LED is lit when the pressure is at the set level 2 Yellow LED is lit when pressure is below the set level Check connections AtmosAir MIN POWER LOW PRESSURE NORMAL PRESSURE OFF ON MAX ALT PR...

Страница 16: ...tation The level of alternating pressure or rotation can be adjusted for patient comfort Prior to engaging alternating pressure or rotation ensure that bed frame has side rails and that all side rails...

Страница 17: ...ransport there are two options 1 Disconnect the tube assembly from the mattress Valves will close off as soon as the tube assembly is disconnected 2 Disconnect the tube assembly from the pump and secu...

Страница 18: ...ss with the fire barrier exposed to avoid contact with loose particles To ensure optimal product performance avoid over handling of the fire barrier and handle with care Fire barrier should be replace...

Страница 19: ...ls 1 Ensure pump is unplugged 2 Wipe surface of pump and tubes with coarse cloth using an approved disinfectant germicide mixed to manufacturer s instructions Do not flood any part of the pump with cl...

Страница 20: ...and hoses of A and AR models should be cleaned weekly Inspection System Check Out Check each of the following before placing the AtmosAir MRS with a new patient 1 Check mattress surface for tears or...

Страница 21: ...ed hoses are connected to hoses marked with red tag for rotation Tubing kinked Check tubing inside mattress for kinks Tubing disconnected Check tubing inside mattress for possible disconnect Hole in r...

Страница 22: ...rotation bladders for A and AR models are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems RF Welded Mattress Seperate Co...

Страница 23: ...and AR models are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems Foam SAT Cover part of SAT system SAT Interliner part...

Страница 24: ...dels are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of SAT Systems Foam SAT Interliner part of SAT system image for reference only...

Страница 25: ...t fits over the foam and SAT assemblies The SAT Cover SAT Cells SAT Interliner are all part of the integrated SAT System and cannot be ordered separately See Replacement Parts for a complete list of S...

Страница 26: ...22 Parts Diagram AtmosAir Pump A and AR models See Replacement Parts for a complete list of parts that may be ordered for the AtmosAir Pump as well as entire pump systems...

Страница 27: ...00 9000 Foam Shell KA4 9FS SIZE APod 25 Foam Shell 25PFS SIZE T Series Foam Shell KA9FSTC V Series Foam Shell KVFS3584 M Series Foam Shell 416119 M Series Orthopedic Foam Shell 416397 AtmosAir with SA...

Страница 28: ...2 Top Vinyl Gray Bottom A9CRLVG WWWLL TXS Complete Cover Barrier Flex 1 Top Non Skid Bottom A9CBFNB WWWLL PXS Complete Cover Barrier Flex 1 Top Non Skid Bottom No Handles A9CBFNH WWWLL PXS Complete Co...

Страница 29: ...n Skid PU Bottom With Handles ADCPRPU WWWLL NXS Complete Cover Soflux Top Non Skid Bottom ADCSONB WWWLL NXS Complete Cover Soflux Top Vinyl Gray Bottom ADCSOVG WWWLL NXS Complete Cover Reliant IS2 Top...

Страница 30: ...ernating Pressure Complete Cover Reliant IS2 top and Non Skid base TXCRLNB34079TXS AtmosAir V Series PART TYPE CATALOG NUMBER V Series Mattress Complete Cover Soflux top and Vinyl Gray base AVCSOVG350...

Страница 31: ...Length 75 in 190 5 cm Maximum Length 84 in 213 cm Minimum Width 32 in 81 5 cm Maximum Width 38 in 96 5 cm Height 7 in 17 5 cm Pump 120V Weight 5 lb 2 5 kg Dimensions 10 25 in 26 cm L x 4 5 in 11 5 cm...

Страница 32: ...to the EMC information in the following tables The following tables document compliance levels and guidance from the IEC 60601 1 2 2007 Standard for the electromagnetic environment in which the Atmos...

Страница 33: ...rior to application of the test level Recommended separation distances between portable and mobile RF communications equipment and the AtmosAir MRS The AtmosAir MRS is intended for use in an electroma...

Страница 34: ...liance level in each frequency range see note b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOT...

Страница 35: ...Foot End Possible electrical shock hazard Warning of possible hazard to system patient or staff Attention See User Guide Consult User Guide On Power Off Manufacturer CE marking indicating conformity...

Страница 36: ...oxygen or nitrous oxide Serial Number The operator must read this document this instruction for use before use Note the symbol is blue on the product label Customer Contact Information For questions r...

Страница 37: ...33 Intentionally left blank...

Страница 38: ...34 Intentionally left blank...

Страница 39: ...mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Office 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 48...

Страница 40: ...ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to conti...

Отзывы: