MAINTENANCE
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
33 / 48
2. The pump generates a pressure
build-up automatically.
When the pressure 0.2 bar reaches
press “START/STOP”.
3. Press "MODE".
Display shows “CAL 1.2b”.
4. The pump continues on with the
pressure build-up.
When the pressure 1.2 bar reaches
press “START/STOP”.
5. Release the pressure and press
"MODE". Display shows “CAL -1.0b”.
6.3.7. Calibrating the downstream sensor
Remark:
The pump must reach a pressure of at least 1.2 bar. Otherwise there
might be a mechanical problem (Check the pump unit and the door).
1. Simulate a downstream occlusion by
the 3-way stop cock.
6. Close the roller clamp and open the
door. The pump pumps reverse.
“CAL 0.2b”
“CAL 1.2b”