SAFETY STANDARD CHECK
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
47 / 48
10. SAFETY STANDARD CHECK
Safety Standard Check (SSC)
ARGUS 707 V_en
Serial-no.: …………………………....……………….....
Inventory-no.: ………………….…..…………….....
Hospital: ………………………………………………….
Department: …………………………………….…..
Customer: ………………………………………………………………………………………………….
The SSC has to be performed at least every 24 months or after 10'000 hours of operation.
The check has to be done in accordance to the user- and service manuals.
1
Check if a software upgrade is required?
2
Visual check for damage, cleanness and com-
pleteness
- Housing, labels, accessories, connectors,
power cable, etc.
3
Test the function of the stop flow clamp
- Proper movement of the clamp
-
Display pump type and software release
-
Display of 2, 4, 7, F., in numeric display
4
Keep "MODE" pressed while switching on the
pump
-
Display of all operation- and alarm indicators
5
Connect/disconnect the pump to the mains
-
The indicator “external supply” turns on/off
6
Test the drop detector by simulating drops
-
Check the green “drop” indicator
-
Door open the “door” indicator lights up
7
Test the door switch, open and close the door
-
Door close the “door” indicator turns off
8
Open the door and remove any IV set
-
Indicator “air bubble” lights up
9
Install a water filled IV set, close the door
-
Indicator “air bubble” turns off
-
The red indicator “drop” lights up (delayed)
10
Set rate to 333.3 ml/h, press "START",
disconnect the drop detector
-
The acoustical alarm turns on
11
Press "MODE"
-
The acoustical alarm mutes
12
Check the external connector “nurse call”
-
Relay contact switches (see
chapter 9
)
Calibration of the pressure sensors
IV set type used:
13
Codan …………...……. Other ………………….
-
See
chapter 6.3.
and
6.4.
of the service
manual
14
Volume calibration
-
See
chapter 6.5.
of the service manual
15
Charge the battery while the pump is running 16
hours, at a rate of 30.0 ml/h
- The indicator “external supply” must light
Battery check at a rate of 30 ml/h.
16
Run the battery test until the pump switches off
automatically
- The green indicator "battery" lights up during
this test
- Check the pump internal time and data
17
Printout the pump history (refer to
chapter 4.4.5.
)
- Check the battery run time by checking the
latest history entries, > 4h 30min
Electrical test according to EN60601-1
- Visual check of mains connector
18
(all measurements made with a power cable 2.5m)
- Measurements attached
19
Charge the battery after this test!
The pump has passed the SSC and is safe for use
Date / Name: ……...……………………………………..
Signature: ………………………………............…
ARGUS Medical AG