![Argus 707 V Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/argus/707-v/707-v_service-manual_2969492011.webp)
PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
10 / 48
2.7. Index list of the pump configuration (with ARGUS service)
Index PC Add. Default
Function
Unit
Range
1
Yes
Run indication by running decimal point
-
No / Yes
2
1
No
Key ON/OFF only at stop valid
-
No / Yes
3
No
Rate change allowed only at STOP
-
No / Yes
4
No
Key STOP delayed
# 361
-
No / Yes
5
No
Second entry of rate
# 3=Yes
-
No / Yes
6
No
Static alarm (staff alerting system)
-
No / Yes
7
No
Display elapsed time in run mode
# 8=No
-
No / Yes
8
No
Display remaining time
# 7=Yes
-
No / Yes
9
10
2
No
Operation without drop detector
-
No / Yes
11
3
Yes
Recall of ml/h rate at next power on
-
No / Yes
12
No
Recall of ml total at next power on
-
No / Yes
13
No
Recall of ml inf. at next power on
-
No / Yes
14
No
SBS (step by step function)
-
No / Yes
15
No
Display VTBI (volume to be infused)
-
No / Yes
16
No
Display “Set-X” if only one set is enabled
-
No / Yes
17
Yes
KVO (KOR) enabled
# 60
-
No / Yes
18
No
Drop alarm only if bottle empty
# 10=No,
→
trA
-
No / Yes
19
4
No
Buzzer at start
-
No / Yes
20
No
Menu Clr (clear ml inf.)
# 15=No
-
No / Yes
21
22
No
Menu trA (transport)
# 10, 18=No
-
No / Yes
23
Yes
Menu PrL (pressure alarm limit)
-
No / Yes
24
Yes
Menu CAP (battery capacity)
-
No / Yes
25
No
Menu SEt Fill (prime IV set)
26
No
Menu InF (ml inf. since last power on)
-
No / Yes
27
No
Menu dLo (data lock)
-
No / Yes
28
No
Menu Stb (stand by)
-
No / Yes
29
No
Menu Med (medication name)
-
No / Yes
30
No
Menu tM (timer alarm)
-
No / Yes
31
32
No
Menu boL (bolus release)
-
No / Yes
33
No
Menu boLr (bolus rate)
# 32=Yes
-
No / Yes
34
No
Menu tot (bolus total)
# 32=Yes
-
No / Yes
35
36
37
38
No
Automatic bolus application
# 32, 34=Yes
-
No / Yes
39
No
Neonatology mode accessible in PrL menu
-
No / Yes
40
41
No
Clear ml/h after infusion completed
-
No / Yes
42
No
Clear ml total after infusion completed
# 41=Yes
-
No / Yes
43
No
Air volume accumulated
-
No / Yes
44
5
Yes
Automatic pressure release after occlusion
-
No / Yes
45
Yes
Pressure display ON (LED bar graph - 20/40/60/80/100%)
-
No / Yes
46
No
Pressure display with indicator
# 45=Yes
-
No / Yes
47
No
Stand by- and battery pre alarm low volume
-
No / Yes
48
No
Flashing numeric display at alarm
-
No / Yes
49
6
No
Alarm acknowledge only with key MODE
-
No / Yes