4.2.2. How to edit a configuration
The following procedure describes how to edit a pump configuration:
1. Press the green “Edit” button.
2. The software will ask for the pump PIN code as next. The button “Edit” changes
its colour and will be renamed into “Download”.
3. If you want to import a configuration from a file press the “Import” button, other-
wise skip this point.
4. Select “Configuration” in the structure tree in the left upper frame.
5. Select the index group you want to modify by selecting the according structure
tree and the according index.
6. Modify the according index (within the given restrictions shown).
Each value (number) must be acknowledged by the green “Enter” button.
Go through point 5 & 6 for all further indexes you want to modify.
7. Press the “Download” button if you want to save the modified configuration on the
pump. Otherwise you can save the modified configuration into a file by pressing
the “Save” button.
8. Make a functional check on each pump you have configured.
Important remark!
If a config. has been edited (performed point 1 and 2) once do not switch off the
pump! Otherwise the pump will change always into the PC-configuration mode
automatically.
4.3. ARGUS service - Calibration (volumetric pums only)
Use this feature to easily calibrate the
volume delivered by the volumetric pump.
To do this, it is possible to enter
the measured rate (of an infu-
sion device tester e.g.) or from
a measured volume (of a bal-
ance) to define a correction fac-
tor according
chapter 6.5.2.