SOFTWARE UPDATES
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
28 / 48
8. Switch the pump ON while keeping key
[10] pressed. Start the “ARGUS service”
tool and select the according COM port.
9. Import
configuration
from backup
. The
calibration values and configuration of last
connected pump will be imported.
10. Download it to the pump by pressing the
“Download” button.
11.
Important:
Perform a standard safety check (
see
chap.10
), the calibration values maybe
lost during the upgrade procedure!
5.4. Safety aspects
Be aware of the following points:
!
For medical device traceability your local distributor or ARGUS Medical AG
needs to be informed about every device updates (serial number) you performed!
!
Do not make any software updates when the device is used and/or connected to
a patient!
CAUTION!
A standard safety check
(see chap.10)
has to be performed after every software
update!