PUMP CONFIGURATIONS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
9 / 48
2.6. Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be
performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS
service
”.
Address
Default
Function
Unit
Range
left display
Index
PC
right display
1
2
No
Key ON/OFF only at stop valid
-
0=No / 1=Yes
2
10
No
Operation without drop detector
-
0=No / 1=Yes
3
11
No
Recall of the last used ml/h rate
-
0=No / 1=Yes
4
19
No
Buzzer at start
-
0=No / 1=Yes
5
44
Yes
Automatic pressure release
-
0=No / 1=Yes
6
49
No
Alarm acknowledge with key MODE
-
0=No / 1=Yes
100
361
5
Key ON/OFF delay time
• 0.1 s
0 - 31
101
362
8
Display brightness
level
2 - 15
102
363
7
Buzzer volume
level
5 - 10
103
316
9
Default pressure limit (IV-set 1)
• 100 mbar
1 - 10
104
318
20
Drop rate window centre (IV-set 1)
Drops/ml
10 - 65
200
317
250
Air bubble size (IV-set 1)
µl
50 - 1000
399
-
707
Enter the calibration menu (volume)
-
123
-
707
Enter the calibration menu (pressure)
-
1234
Note!
The address does not correspond with the index used by the “ARGUS
service
” tool.