MAINTENANCE
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
32 / 48
5. Release the pressure and press
"MODE".
Display shows “uPSt rEAM”.
3. Open the roller clamp, insert the
new IV set and close the door.
1. Take a new IV set filled with water.
1. Install a clamping shears near the
pump (bottle side) and wait 15 sec,
then press “START/STOP”. *
2. Release the clamping shears and
press “MODE”.
Display shows “CAL 0.2b”.
6.3.5. Calibrating the downstream sensor
6.3.6. Calibrating the upstream sensor part 1
*
Remark:
By pressing key “1” the pressure signal in (mV) of the upstream sensor
will be displayed, after approximately 15 seconds the value does not
decrease that fast any more; at this point press “START/STOP” to store
the value (refer to step 1). The software will not allow the value to drop
lower than the previous saved door open offset value +50mV.
2. Simulate a downstream occlusion
by the 3-way stop cock before you
insert the set.
4. The pump generates a pressure
build-up automatically.
When the pressure 1.0 bar
reaches press “START/STOP”.
“uPSt rEAM”
“CAL. 1.0b”