ARGUS service
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
19 / 48
4. ARGUS
SERVICE
4.1. General
The new ARGUS
service
utility is a high and user friendly PC software which can
configure and upgrade pumps over PC serial COM port. With this Windows based
software you can also set pump clock, change PIN code, read and print out history
and easily replicate pump configurations, and so on. The modern and clearly struc-
tured design of this self-describing PC-tool allows a very easy and rapid modifica-
tion of the ARGUS 600 Syringe pump, the ARGUS 707 & 708 Volumetric pump and
the ARGUS docking station. This software may be available from your local distribu-
tor or directly from ARGUS.
REMARK:
“ARGUS
service
” may only be used with software versions greater or equal to 4.00.
CAUTION!
The infusion pump has to be disconnected from the patient before and while the se-
rial interface cable is connected to the pump.
The connection of the A707 over the serial interface RS-232 can be done by con-
necting the interface cable (
REF
10.093) to the serial interface outlet of the serial
PC-COM port.
Start the ARGUS
service
Press the button which confirm to your
previously connected device
(ARGUS pump or docking station).
Hold the “10” key while switching-ON the
pump.
Select “Start Configuration” (
see next
chapter
) or “Start Update-Center” (
see
chapter 5.3.2.
).